CSlaser LT

Infrared thermometer optris CSlaser LT

The infrared thermometer optris CSlaser LT consists of a robust stainless steel sensor with integrated electronics, enabling easy integration into your system.

Its innovative double laser sight allows for an exact marking of measurement spots even on smallest objects. A variety of optics enables adaptation for a broad range of different applications. Additionally, the IR thermometer offers a standardized two-wire interface for a reliable transfer of measurement data and easy integration into a Siemens SPS.

Important specifications

  • Temperature range:
    -30°C to 1,000°C (-22 °F to 1832 °F)
  • Spectral range: 8 to 14 µm
  • Response time: 150 ms
  • Starting at $635

From $705.00

Technical Details of the optris CSlaser LT

Measurement Specifications

  • Temperature range (scalable via software): -30°C … 1,000°C (-22 °F … 1832 °F)
  • Spectral range: 8 to 14 µm
  • Optical resolution (90% energy): 50:1
  • System accuracy (at Tamb = 23 ± 5°C [73.4 ±41 °F]): ± 1% or ± 1°C (±1.8 °F) 1)
  • Repeatability (at Tamb = 23 ± 5°C [73.4 ±41 °F]): ± 0.5% or ± 0.5°C (±0.9 °F) 1)
  • Temperature resolution: 0.1 K
  • Response time (90% signal): 150 ms
  • Emissivity / Gain (adjustable via sensor or software): 0.100 – 1.100
  • IR window correction (adjustable via software): 0.100 – 1.100
  • Signal processing (parameter adjustable via software):
    peak hold, valley hold, average, extended hold function with threshold and hysteresis

General Specifications

  • Environmental rating: IP 65 (NEMA-4) front mountable at vacuum processes (up to 10-3 mbar)
  • Ambient temperature: -20°C … 85°C (-4 °F … 185 °F) (50°C [122 °F] with laser ON)
  • Storage temperature: -40°C … 85°C (-40 °F … 185 °F)
  • Relative humidity: 10 – 95%, non condensing
  • Vibration: IEC 60068-2-6 (sinus shaped), IEC 60068-2-64 (broadband noise)
  • Shock: IEC 60068-2-27 (25G and 50G)
  • Weight: 600 g (21.7 oz)

Electrical Specifications

  • Output / analog: 4-20 mA
  • Loop resistance: max. 1000 Ω2)
  • Output / alarm: 0-30 V/500 mA (open collector)
  • Output/ digital: uni / bidirectional, 9.6 kBaud, 0/3 V digital level / USB optional
  • Cable length: 3 m [9.8 ft], 8 m [26.2 ft], 15 m [49.2 ft]
  • Power supply: 5-30 V DC
  • Current draw:
    45 mA at 5 V
    20 mA at 12 V
    12 mA at 24 V

Versions

Part number: temperature range / spectral range / optics

  • OPTCSLLT: -30-1000°C (-22 … 1832 °F) / 8-14 μm / 50:1

1) whichever is greater
2) in dependence on supply voltage